Common problems in the selection of Royreal wet gas flow meter

Time:2024-05-05

c4c0a822-87a2-4865-926a-eb2d822c6199.jpg

Q: What is the working pressure range of ROYREAL wet gas flow meter?

A: The ROYREAL wet gas flow meter is suitable for operating pressures from -10kPa to 10kPa, with custom service available if higher operating pressures are required.

 

Q: The ROYREAL wet gas flow meter operates at an ambient temperature of 18 °C to 22 °C, is the meter only suitable for this ambient temperature?

A: No, the reason why the wet gas meter is recommended to be used at 18 °C to 22 °C is that the test data (quoted data) of the calibration standards and international calibration standards in various regions are tested at the standard temperature of 20 °C, including the calculation formula is also 20 °C, and the test data within this temperature range can be recognized by various laboratories.

The wet gas meter can be used between -10℃ and 50℃, the laboratory environment temperature will not affect the work of the meter, but the test data and 18 °C to 22 °C test data there is a certain deviation. (Due to the principle of thermal expansion and contraction of objects, the amount of liquid filled needs to be adjusted at different temperatures.)